Interference of bovine hemoglobin-based oxygen carrier-201 (Hemopure) on four hematology analyzers
Marinke W Bronkhorst-van der Helm1, Floor Weerkamp2, Albert Huisman3
1Department of Clinical Chemistry, Erasmus MC, University Medical Center, Rotterdam, The Netherlands.
Introduction:
Hemoglobin-based oxygen carriers, for example HBOC-201 (Hemopure), are aimed to bridge acute anemia when blood transfusion is not available or refused by the patient. However, since HBOC-201 appears free in plasma, it interferes with laboratory tests. This study presents an overview of HBOC-201 interference on four commonly used hematology analyzers and suggests treatment monitoring possibilities.
Methods:
Blood samples were spiked with therapeutic doses of HBOC-201 and nine hematology parameters were measured with the Sysmex XN-20, Siemens Advia 2120i, Abbott Alinity Hq and Abbot Cell Dyn Sapphire hematology analyzers. The results were compared to control samples and the bias was determined.
Results:
Most parameters, including all cell counts, hematocrit and MCV, showed a non-significant bias compared to control. However, the standard, total hemoglobin (Hb) measurement as well as MCH and MCHC showed poor agreement with control, as HBOC-201 was included in this measurement. Yet, the flow cytometry-based Hb method quantified intracellular Hb in spiked samples, excluding HBOC-201.
Conclusion:
Of all included hematology parameters, only total Hb and the associated MCH and MCHC suffered from interference. In contrast, the flow cytometry-based Hb measurement provided an accurate measure of intracellular Hb. The difference between total Hb and cellular Hb represents the HBOC-201 concentration and can be used to monitor HBOC-201 treatment.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Blood Transfusion and Agglutination
History
The history of blood transfusion dates back to the 17th century, when early attempts were made in animals. In 1818 James Blundell, a British doctor, performed the first successful human blood transfusion. Later in 1900, Karl...
Blood Transfusion
Blood Transfusion Overview
A blood transfusion is a medical procedure used to replace blood lost due to injury, surgery, or to treat conditions such as anemia or cancer. During a transfusion, donor blood is...
Oxygen Transport in the Blood
Hemoglobin
When all four heme groups are bound to oxygen, the resulting molecule is called oxyhemoglobin. As a result, arterial blood...