Commutability evaluation of candidate reference materials and ERM-DA470k/IFCC for immunoglobulin M using two
Lu Sang1, Chunying Wu2, Huijuan Chen1
1Department of Clinical Laboratory, Beijing Huairou Hospital, Beijing, China.
Background:
This study aimed to assess the commutability of frozen pooled human serum (PHS), high concentration of Immunoglobulin M (IgM) pure diluted materials (HPDM), commercialized pure materials (CPM), and dilutions of ERM-DA470k/IFCC in IgM detection using the CLSI and IFCC approaches, to support standardization or harmonization of IgM measurement.
Methods:
Twenty-four serum samples, relevant reference materials (PHS, HPDM, CPM), and different ERM-DA470k/IFCC dilutions were analyzed in triplicate using six routine methods. The commutability of the relevant reference materials was carried out following CLSI EP30-A and IFCC bias analysis.
Results:
According to the CLSI approach, low, medium, and high concentrations of PHS, HPDM, and CPM were commutable on 10, 13, 15, 13, and 8 of 15 assay combinations, respectively. Using the IFCC approach, low, medium, and high concentrations of PHS, HPDM, and CPM were commutable on 10, 11, 9, 15, and 10 of 15 assay combinations, respectively. The ERM-DA470k/IFCC dilutions with D-PBS and RPMI-1640 Medium were commutable on 13 of 15 assay combinations according to CLSI and were commutable on all 15 assay combinations using IFCC approach.
Conclusions:
High concentration of PHS were commutable on all six detection systems using the CLSI approach. Low and medium concentration of PHS showed unsatisfied commutability. HPDM, not CPM have good commutability, has the potential to become reference materials. ERM-DA470k/IFCC diluted with different medium showed different commutability.