Evaluation of the Ethos synthetic computed tomography for bolus-covered surfaces
Sonja Wegener1, Robert Schindhelm1, Jörg Tamihardja1
1University Hospital Wurzburg, Department of Radiation Oncology, Josef-Schneider-Str. 11, 97080 Wuerzburg, Germany.
Purpose:
Ethos allows online adaption of radiotherapy treatment plans. Dose is calculated on synthetic computed tomographies (sCT), CT-like images generated by deforming planning CTs (pCT) onto daily cone beam CTs (CBCT) acquired during treatment sessions. Errors in sCT density distribution may lead to dose calculation errors. sCT correctness was investigated for bolus-covered surfaces.
Methods:
pCTs were recorded of a slab phantom covered with bolus of different thicknesses and with air gaps introduced by spacer rings of variable diameters and heights. Treatment plans were irradiated following the adaptive workflow with different bolus configurations present in the pCT and CBCT. sCT densities were compared to those of the pCT for the same air gap size. Additionally, the neck region of an anthropomorphic phantom was imaged using a plane standard bolus versus an individual bolus adapted to the phantom's outer contour.
Results:
Varying bolus thickness by 5 mm between pCT and CBCT was reproduced in the sCT within 2 mm accuracy. Different air gaps in pCT and CBCT resulted in highly variable bolus thickness in the sCT with a typical error of 5 mm or more. In extreme cases, air gaps were filled with bolus material density in the sCT or the phantom was unrealistically deformed near changed bolus geometries. Changes in bolus thickness and deformation also occurred in the anthropomorphic phantom.
Conclusion:
sCTs must be critically examined and included in plan-specific quality assurance. The use of tight-fitting air gap-free bolus should be preferred to increase the similarity between sCT and CBCT.
More Related Videos
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Ultrasonography
During an ultrasonography procedure, a handheld device called...