Comparison of the performance of two automatic cell morphology analyzers for peripheral-blood leukocyte morphology
Shihong Zhang1, Yongjian He2, Wenjing Wu3
1Department of Laboratory Medicine, The First Affiliated Hospital of Sun Yat-sen University, Guangzhou, China.
Introduction:
To compare the morphological classification ability of peripheral-blood leukocytes of the automatic cell morphology analyzers MC-100i and DI-60.
Methods:
(1) MC-100i and DI-60 were used to analyze leukocytes in 432 venous blood samples collected from three tertiary hospitals across China. The preclassification results were compared with the results reported by senior morphological experts (postclassification results) to evaluate the accuracy, sensitivity, specificity, and consistency of leukocyte preclassification for both instruments. (2) In 200 of the 432 blood samples, morphological experts conducted manual microscopic examination for various types of leukocytes. The correlation between the MC-100i and DI-60 leukocyte postclassification results and the expert microscopist results were analyzed.
Results:
(1) MC-100i preclassified leukocytes and nucleated red blood cells (RBCs). Compared with the postclassification results, the total leukocyte preclassification accuracy of MC-100i was 97.16%, while that of DI-60 was 87.24%. The sensitivity of MC-100i to abnormal cells (including blasts, promyelocytes, neutrophilic myelocytes, neutrophilic metamyelocytes, reactive lymphocytes, abnormal promyelocytes, plasma cells, abnormal lymphocytes and nucleated RBCs) was 90.24%, which was significantly higher than the 50.72% sensitivity of DI-60. (2) Comparing the postclassification results with manual microscopy, except for reactive lymphocytes and basophils, the MC-100i and DI-60 results had good correlations with various leukocyte types and nucleated RBCs (r > 0.85), and MC-100i was better than DI-60 in the recognition of basophils.
Conclusion:
Both MC-100i and DI-60 have good detection ability for five normal types of leukocytes in peripheral blood. MC-100i has significantly better detection sensitivity for abnormal cells in peripheral blood than DI-60.
More Related Videos
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...