Quantitative injection strip platform using water-soluble paper and magnet based on a lateral flow assay
Yewon Kwon1, Dami Kim2, Sanghyo Kim1,2
1Department of Bionanotechnology, Gachon University, Seongnam 13120, Republic of Korea. rnjsdpdnjs@gachon.ac.kr.
A novel quantitative injection (QI) strip ensures consistent sample volume for precise lateral flow immunoassay (LFA) results. This innovation reduces variability, making LFA strips more reliable for quantitative analysis without extra equipment.
Area of Science:
- Biomedical Engineering
- Analytical Chemistry
- Assay Development
Background:
- Quantitative analysis in lateral flow immunoassay (LFA) is crucial for accurate concentration measurement.
- Traditional LFA strips can suffer from signal variability due to inconsistent sample volumes.
- Improving the precision of LFA strips is essential for point-of-care diagnostics.
Purpose of the Study:
- To design and evaluate a novel quantitative injection (QI) strip for improved LFA accuracy.
- To address the challenge of variable sample volumes in conventional LFA devices.
- To enhance the reliability of strip-based quantitative analysis.
Main Methods:
- Modification of LFA strip structure using water-soluble paper and a magnet to ensure a consistent unit volume injection.
- Incorporation of a physical separation mechanism for the gold conjugate pad and nitrocellulose membrane at an optimized time.
- Comparative analysis of detection signal consistency and deviation between QI strips and standard LFA strips across varying sample volumes.
Main Results:
- The QI strip successfully maintained consistent detection signals regardless of injected sample volume.
- Standard LFA strips showed increased detection signals with larger sample volumes.
- The QI strip exhibited an 11-fold reduction in data deviation compared to standard LFA strips.
- The integrated separation function was identified as the key factor for improved performance.
Conclusions:
- The QI strip offers a user-friendly and equipment-free solution for precise quantitative analysis in LFAs.
- QI strips demonstrate superior suitability for quantitative applications over traditional LFA strips.
- This technology holds potential for advancing user-friendly point-of-care testing (POCT) and strip-based diagnostics.
More Related Videos
10:10Development of a Lateral Flow Immunochromatographic Strip for Rapid and Quantitative Detection of Small Molecule Compounds
Published on: November 13, 2021
08:22Electrowetting-based Digital Microfluidics Platform for Automated Enzyme-linked Immunosorbent Assay
Published on: February 23, 2020