Optimal dose determination for coronary artery calcium scoring CT at standard tube voltage
Yifang Zhou1, Emi Eastman1, Christina Lee1
1Department of Imaging, Cedars-Sinai Medical Center, 8700 Beverly Blvd., Los Angeles, CA 90048, USA.
Objectives:
Coronary artery calcium scoring (CACs) at 120 kVp is the standard practice. It is an important tool for preventative management of asymptomatic patients. However, the current dose delivery, albeit patient-size dependent, does not connect the CACs specific noise requirement to the dose, causing significant dose variations. We propose a new approach for optimal dose determination by incorporating the patient-size dependent noise threshold.
Methods:
A polyethylene-based Mercury phantom of various diameters was scanned with a dual-source CT using CACs gating at different volume CT dose index (CTDIvol). The relationship of noise to the diameter and CTDIvol was obtained. The phantom diameter was then converted to the patient chest diameter through a retrospective analysis of a clinical cohort (N = 140). Finally, the patient-size dependent noise threshold was applied, and the optimal dose was derived. The prescribed doses were compared with those from a clinical CACs cohort (N = 262).
Results:
A power-exponential relationship was found for the noise versus CTDIvol and phantom diameter (R2 = 0.988). The phantom diameter versus the patient effective diameter was found to obey a linear relationship (R2 = 0.998). Two noise threshold settings were made for dose options: one for more dose saving, and another for tighter noise constraint. Retrospective comparisons with clinical CACs studies showed an average dose reduction of 23% in 80.5% of the cases with option 1. The average dose reduction is 23% in 77.9% of the cases with option 2.
Conclusion:
A new optimal dose scheme dictated by the target noise was established for CACs at 120 kVp. The proposed dose modulation can serve as the baseline from which further dose reduction is possible.
More Related Videos
06:57Author Spotlight: Advancing Cardiovascular Imaging - Introducing the Spatially Weighted Calcium Score for Early Disease Detection
Published on: September 22, 2023
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography