Establishing local diagnostic reference levels for computed tomography examinations using size-specific dose
Faisal A Alrehily1, Khalid S Alqahtani1, Marwan H Aljohani1
1From the Department of Diagnostic Radiology Technology (Alrehily, Alqahtani, Aljohani, Alharbi, Abdulaal, Alshamrani, Alhazmi), College of Applied Medical Sciences, Taibah University; from the Department of Radiology (Alharbi), King Salman bin Abdulaziz Medical City; from the Department of Nuclear Medicine (Alsaedi), King Fahad General Hospital (Alsaedi), Al Madinah Al Munawwarah, Kingdom of Saudi Arabia; from the Department of Radiological Techniques (Al-Murshedi,), AL-Zahraa University for Women, College of Health and Medical Technology, Karbala; and from Physics Department (Al-Murshedi), College of Education for Pure Sciences, University of Babylon, Iraq.
Objectives:
To establish local DRL (LDRL) for computed tomography (CT) examinations based on size-specific dose estimates (SSDEs), which consider patient size. The concept of diagnostic reference level (DRL) was introduced to limit patient exposure to unnecessary radiation. However, traditional DRL values do not consider patient size.
Methods:
Following institutional committee approval, data were collected from CT examinations of adult patients at Madinah General Hospital, Al Madinah Al Munawwarah, Saudi Arabia from January to March 2023. The SSDE was calculated for each patient using the effective diameter (Deff).
Results:
The LDRLs of the brain, cervical spine, chest, thoracic spine and kidneys, ureters, and bladder (KUB) examinations were 118 mGy, 12 mGy, 8 mGy, 17 mGy, and 7 mGy, respectively. A strong correlation was observed between SSDEs and the volume computed tomography dose index (CTDIvol) for all examinations except chest scans (p<0.05). Size-specific dose estimates were higher than the CTDIvol, with a greater difference for patients with smaller Deff (p<0.05).
Conclusion:
The established LDRL was within the international DRL. The use of SSDE has the potential to provide more accurate and relevant data for radiation safety practices; however, widespread adoption of SSDE in new CT scanners is necessary for promoting consistency and standardization methodologies.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT