Related Experiment Video
Updated: Jul 19, 2025

05:41
Author Spotlight: Integrating Ultrasound Imaging with Biochemical Markers for Thyroid Disease Diagnosis
Published on: February 9, 2024
663
Differentiating thyroid nodules parathyroid lesions using 2D-shear-wave elastography: a novel approach for enhanced
Dana Stoian1,2, Andreea Borlea1,2, Laura Taban3
1Discipline of Endocrinology, Second Department of Internal Medicine, University of Medicine and Pharmacy "Victor Babes", Timisoara, Romania.
Frontiers in Endocrinology
|August 17, 2023
Summary
Bidimensional shear wave elastography planewave ultrasound (2D SWE PLUS) effectively distinguishes parathyroid lesions from thyroid nodules. This advanced ultrasound technique improves diagnostic accuracy for superficial neck structures.
Area of Science:
- Medical Imaging
- Endocrinology
- Oncology
Background:
- Distinguishing thyroid and parathyroid lesions using conventional ultrasound can be challenging.
- Accurate differentiation is crucial for appropriate patient management and treatment planning.
Purpose of the Study:
- To evaluate the diagnostic efficacy of bidimensional shear wave elastography planewave ultrasound (2D SWE PLUS) in differentiating thyroid nodules from parathyroid lesions.
- To assess 2D SWE PLUS as an auxiliary tool to enhance conventional ultrasonography for superficial neck structures.
Main Methods:
- Eighty-six patients with concurrent thyroid nodules and hyperparathyroidism were evaluated.
- Conventional ultrasound and 2D SWE PLUS were performed using an Aixplorer Supersonic Mach30 with a 5-18 MHz linear probe.
- Elasticity index (EI) values were statistically analyzed to compare parathyroid tissue, normal thyroid tissue, and thyroid nodules.
Main Results:
- Significant differences in EI were found between parathyroid and normal thyroid tissue (p<0.0001), and between parathyroid lesions and thyroid nodules (p<0.0001).
- 2D SWE PLUS achieved high diagnostic performance with an area under the curve (AUC) of 0.961 for parathyroid vs. normal thyroid tissue (cut-off ≤8.9 kPa) and 0.963 for parathyroid vs. thyroid lesions (cut-off 9.24 kPa).
- Elevated elasticity values were observed in thyroid cancer compared to benign thyroid nodules (p<0.0001).
Conclusions:
- 2D SWE PLUS is a highly effective tool for differentiating thyroid nodules from parathyroid lesions.
- This technique significantly enhances diagnostic performance in neck ultrasonography, aiding in the distinction of these superficial structures.
- The findings support the integration of 2D SWE PLUS into routine ultrasonography for improved diagnostic accuracy in endocrine neck imaging.