Quantitative study on dose distribution of Freiburg flap for keloid high-dose-rate brachytherapy based on MatriXX
Jie Ni1, Guanghui Gan1, Xiaoting Xu1
1Radiation Therapy Center, The First Affiliated Hospital of Soochow University, Suzhou, Jiangsu Province, China.
Purpose:
To quantify the dose distribution effect of insufficient scattering conditions in keloid HDR brachytherapy with Freiburg fFlap (FF) applicator.
Materials And Methods:
A phantom composed of FF applicator, MatriXX and solid water slices was designed and scanned for treatment planning. Bolus with different thicknesses were covered to offer different scatter conditions. Planar dose distributions were measured by MatriXX. The maximum value (Max), average value (Avg) and γ passing rate (3 mm/3%) were evaluated by the software MyQA Platform.
Results:
The maximum and average doses measured by MatriXX were lower than the calculated values. The difference increased as field size decreased. The Max value, found at 0.86 cm level in the two tube case, was -20.0%, and the avg value was -11.9%. All the γ values were less than 95%. This difference gradually decreased with increasing bolus thickness and the γ values were significantly improved.
Conclusion:
MatriXX could be used for dose verification of HDR brachytherapy with an FF applicator. When the FF applicator was applied for keloid, insufficient scattering conditions would cause an insufficient target dose. This difference could be reduced by covering the bolus with different thicknesses on the applicator. The smaller the field, the thicker the bolus required.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Rational Dosage Regimen: Maintenance Dose and Loading Dose
In most cases, drugs are administered repetitively or infused continuously to maintain a steady-state concentration in the body. At a steady state,...
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses