Presentation of Organ Dose and Effective Dose Conversion Factors in Dual-Energy Computed Tomography: A Monte Carlo
1Department of Physics, Ahvaz Branch, Islamic Azad University, Ahvaz, Iran.
Background:
The same conversion factors (k-factors) of Single CT (SECT) are applied to estimate the Effective Dose (ED) in Dual Energy Computed Tomography (DECT). However, k-factors for different organs need independently validating for DECT, due to the different conditions in DECT.
Objective:
This study aimed to calculate organ dose and k-factors in different imaging protocols (liver, chest, cardiac, and abdomen) for male and female phantoms.
Material And Methods:
This Monte Carlo Simulation study used Monte Carlo N-Particle (MCNP) code for modeling a Siemens Somatom Definition Flash dual-source CT scanner. The organ dose, dose length product, and k-factors were calculated for the Medical Internal Radiation Dose (MIRD) of male and female phantoms.
Results:
For the male phantom, the k-factors for the liver, chest, cardiac, and abdomen-pelvis imaging protocols are equal to 0.020, 0.012, 0.016, and 0.014 mSv.mGy-1cm-1, respectively. For the female phantom, the corresponding values are equal to 0.026, 0.023, 0.036, and 0.018, respectively. These values for DECT are different from those corresponding values for SECT, especially for the female phantom.
Conclusion:
The calculated k-factors for DECT can be used as reference values for the estimation of ED in DECT.
More Related Videos
Related Concept Videos
Biological Effects of Radiation
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Dose-Response Relationship: Overview