Randomized Trial Comparing Prediction Accuracy of Two Swept Source Optical Coherence Tomography Biometers
Sam Multack1, Nellie Plummer1, Gerard Smits2
1Multack Eye Care, Frankfort, IL, USA.
Clinical Ophthalmology (Auckland, N.Z.)
|August 23, 2023
Summary
The Argos and IOLMaster 700 biometers demonstrated similar refractive prediction accuracy for cataract surgery. While the IOLMaster 700 showed a statistically significant difference in absolute prediction error, it was not clinically meaningful.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Medical Devices
Background:
- Accurate biometry is crucial for successful intraocular lens (IOL) power calculation in cataract surgery.
- Standard keratometry and Total Keratometry measurements can influence refractive outcomes.
- Comparing new biometry devices to established ones is essential for clinical adoption.
Purpose of the Study:
- To compare the refractive prediction accuracy of the Argos biometer (using standard keratometry) with the IOLMaster 700 (using Total Keratometry).
Main Methods:
- A randomized, prospective study involving 80 patients.
- Preoperative biometry was performed using both Argos and IOLMaster 700 devices.
- Prediction errors were calculated based on 1-month postoperative manifest refraction after cataract surgery and IOL implantation.
Main Results:
- Directional prediction errors were similar between devices (Argos: 0.07 ± 0.32 D; IOLMaster 700: 0.08 ± 0.34 D).
- Absolute prediction errors showed a statistically significant difference (Argos: 0.21 ± 0.25 D; IOLMaster 700: 0.25 ± 0.24 D), but this was not clinically significant.
- The percentage of eyes within 0.5 D of absolute prediction error was high for both: 91% for Argos and 88% for IOLMaster 700, with no significant difference.
Conclusions:
- The Argos and IOLMaster 700 biometers exhibit comparable refractive prediction accuracy in eyes with normal axial lengths.
- Despite a statistically significant difference in mean absolute prediction error, the variation between the devices is not clinically meaningful for cataract surgery outcomes.