Non-Invasive Estimation of Central Systolic Blood Pressure by Radial Tonometry: A Simplified Approach
Denis Chemla1, Davide Agnoletti2,3, Mathieu Jozwiak4,5
1INSERM UMRS 999, Hôpital Marie Lannelongue, 92350 Le Plessis-Robinson, France.
Backround:
Central systolic blood pressure (cSBP) provides valuable clinical and physiological information. A recent invasive study showed that cSBP can be reliably estimated from mean (MBP) and diastolic (DBP) blood pressure. In this non-invasive study, we compared cSBP calculated using a Direct Central Blood Pressure estimation (DCBP = MBP2/DBP) with cSBP estimated by radial tonometry.
Methods:
Consecutive patients referred for cardiovascular assessment and prevention were prospectively included. Using applanation tonometry with SphygmoCor device, cSBP was estimated using an inbuilt generalized transfer function derived from radial pressure waveform, which was calibrated to oscillometric brachial SBP and DBP. The time-averaged MBP was calculated from the radial pulse waveform. The minimum acceptable error (DCBP-cSBP) was set at ≤5 (mean) and ≤8 mmHg (SD).
Results:
We included 160 patients (58 years, 54%men). The cSBP was 123.1 ± 18.3 mmHg (range 86-181 mmHg). The (DCBP-cSBP) error was -1.4 ± 4.9 mmHg. There was a linear relationship between cSBP and DCBP (R2 = 0.93). Forty-seven patients (29%) had cSBP values ≥ 130 mmHg, and a DCBP value > 126 mmHg exhibited a sensitivity of 91.5% and specificity of 94.7% in discriminating this threshold (Youden index = 0.86; AUC = 0.965).
Conclusions:
Using the DCBP formula, radial tonometry allows for the robust estimation of cSBP without the need for a generalized transfer function. This finding may have implications for risk stratification.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.