Related Experiment Video
Updated: Aug 13, 2026

How to Administer Near-Infrared Spectroscopy in Critically ill Neonates, Infants, and Children
Published on: August 19, 2020
Severe injury and death associated with home infant cardiorespiratory monitors
Abstract:
Five cases of electrical injury to young children caused by misuse of components of home cardiorespiratory monitors are reported. The injuries, which included one electrocution, occurred when partially or completely disconnected electrode wires were inserted, by an older monitored child or preschool-aged sibling, into a live power cord or an uncovered wall outlet. Anticipatory guidance of home monitor users should emphasize potential electrical injuries and appropriate injury-control behaviors.