Related Experiment Video
Updated: Jul 18, 2025

Optimized System for Cerebral Perfusion Monitoring in the Rat Stroke Model of Intraluminal Middle Cerebral Artery Occlusion
Published on: February 17, 2013
AI-support for the detection of intracranial large vessel occlusions: One-year prospective evaluation
K G van Leeuwen1, M J Becks1, D Grob1
1Department of Medical Imaging, Radboud University Medical Center, Nijmegen, the Netherlands.
Purpose:
Few studies have evaluated real-world performance of radiological AI-tools in clinical practice. Over one-year, we prospectively evaluated the use of AI software to support the detection of intracranial large vessel occlusions (LVO) on CT angiography (CTA).
Method:
Quantitative measures (user log-in attempts, AI standalone performance) and qualitative data (user surveys) were reviewed by a key-user group at three timepoints. A total of 491 CTA studies of 460 patients were included for analysis.
Results:
The overall accuracy of the AI-tool for LVO detection and localization was 87.6%, sensitivity 69.1% and specificity 91.2%. Out of 81 LVOs, 31 of 34 (91%) M1 occlusions were detected correctly, 19 of 38 (50%) M2 occlusions, and 6 of 9 (67%) ICA occlusions. The product was considered user-friendly. The diagnostic confidence of the users for LVO detection remained the same over the year. The last measured net promotor score was -56%. The use of the AI-tool fluctuated over the year with a declining trend.
Conclusions:
Our pragmatic approach of evaluating the AI-tool used in clinical practice, helped us to monitor the usage, to estimate the perceived added value by the users of the AI-tool, and to make an informed decision about the continuation of the use of the AI-tool.
Related Concept Videos
Aneurysm II: Clinical Manifestations and Diagnostic Studies
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies for Cardiovascular System V: CT
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation