Performance evaluation of two different software programs for automated ASPECTS scoring in patients with suspected stroke
Related Experiment Video
Updated: Jul 17, 2025

Computerized Adaptive Testing System of Functional Assessment of Stroke
Published on: January 7, 2019
Carolina Río Bártulos1, Lukas Pirl1, Dennis Lier1
1Insitut für Röntgendiagnostik und Nuklearmedizin, Städtisches Klinikum Braunschweig gGmbH, Braunschweig, Germany.
Background:
The range of software available to radiologists has increased enormously with the advancement of AI. A good example of this is software to determine ASPECTS in the treatment of potential stroke patients.
Objective:
In this study, two software packages (eASPECTS from Brainomix and VIA_ASPECTS from Siemens) were tested and compared for their performance in the daily clinical routine of a maximum care provider with a 24/7 stroke unit.
Methods:
A total of 637 noncontrast CT images were obtained from consecutive patients with suspected stroke, of whom 73 were finally diagnosed with MCA infarction. Differences in agreement and quantification of agreement were analysed, as well as the correlation and sensitivity, specificity and accuracy compared to raters.
Results:
Compared to VIA_ASPECTS, eASPECTS shows good agreement and strong correlation with the raters. VIA_ASPECTS has lower accuracy and low specificity than eASPECTS but a higher sensitivity.
Conclusion:
Both software products have the potential to be decision support tools for radiologists. There are, however, differences between the two software products in terms of their intended use.