Continuous Monitoring of Changes in Heart Rate during the Periprocedural Course of Carotid Artery Stenting Using a

Kentaro Hirai1, Yasunori Fujimoto1, Yohei Bamba1

  • 1Department of Neurosurgery, Osaka Rosai Hospital.

PubMed

Insights

Continuous heart rate (HR) monitoring after carotid artery stenting (CAS) using a wearable device showed a prolonged HR decrease. This finding offers new insights into carotid sinus baroreceptor function post-procedure.

Area of Science:

  • Cardiology
  • Medical Devices
  • Physiology

Background:

  • Carotid artery stenting (CAS) is a procedure to prevent stroke.
  • Heart rate (HR) changes after CAS are not fully understood.
  • Continuous monitoring can provide detailed physiological insights.

Purpose of the Study:

  • To evaluate periprocedural HR changes during CAS using a wearable device.
  • To assess the accuracy of wearable HR monitoring against electrocardiogram (ECG).
  • To investigate the duration and pattern of HR changes post-CAS.

Main Methods:

  • Prospective observational study.
  • Continuous HR monitoring with a wearable device from pre-admission to post-discharge.
  • Bland-Altman analysis to compare wearable device HR with ECG during CAS.
  • Time-series analysis of HR data.

Main Results:

  • A decrease in HR was observed starting on day 1 post-CAS.
  • The most significant HR decrease rate was 12.1% on day 4.
  • Wearable device HR measurement showed good accuracy (bias -1.12 bpm, precision 3.16 bpm) compared to ECG.
  • HR decrease post-CAS persisted longer than previously reported.

Conclusions:

  • Continuous wearable HR monitoring is accurate for assessing periprocedural changes.
  • HR decrease following CAS may be more prolonged than previously thought.
  • Findings suggest sustained effects on carotid sinus baroreceptor physiology.

Related Concept Videos

Pulse rhythm01:30

Pulse rhythm

Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
832
Holter Monitor: 24-Hour Monitoring01:23

Holter Monitor: 24-Hour Monitoring

Holter monitoring is a continuous electrocardiography (ECG) recording that tracks the heart's electrical activity over an extended period, generally 24 to 48 hours. This noninvasive diagnostic tool detects irregular heart rhythms that may not be captured during a standard ECG performed in a clinical setting.DeviceThe Holter monitor is a portable, small device connected to several electrodes on the patient's chest. These electrodes detect the heart's electrical signals and transmit them to the...
52
Assessment of blood pressure in brachial artery(two-step method)01:23

Assessment of blood pressure in brachial artery(two-step method)

Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
757
Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
1.1K