Reliability of Intraocular Pressure Measurements in a Low-Contact Drive-Through Setting
Lynn W Sun1, Yunshu Zhou, Joshua D Stein
1Department of Ophthalmology & Visual Sciences, University of Michigan Kellogg Eye Center, Ann Arbor, MI.
Journal of Glaucoma
|September 6, 2023
Summary
Drive-through iCare tonometry for intraocular pressure (IOP) monitoring may overestimate readings. Confirm significant IOP changes with in-clinic measurements to ensure accurate glaucoma patient care.
Area of Science:
- Ophthalmology
- Glaucoma Management
- Medical Technology
Background:
- Drive-through intraocular pressure (IOP) measurement using iCare tonometry offers low-contact, high-throughput monitoring.
- The iCare tonometer may be susceptible to errors from variable measurement techniques and air currents, potentially leading to IOP overestimation.
Purpose of the Study:
- To assess the reliability of drive-through IOP measurements.
- To compare drive-through IOP measurements with in-clinic measurements during the COVID-19 pandemic for glaucoma patient monitoring.
Main Methods:
- A retrospective study compared drive-through IOP measurements with recent in-clinic measurements.
- Data from 314 patients (868 measurements) were analyzed, excluding those with pre-drive-through IOPs >21 mmHg.
- Measurements were compared to the closest previous and/or subsequent in-clinic IOP readings.
Main Results:
- Drive-through IOP measurements were, on average, 2.4 mmHg higher than in-clinic measurements.
- 68.6% of drive-through IOPs were higher than corresponding in-clinic readings, with 25.9% showing an increase of over 5 mmHg.
- Differences were observed in both the right (OD) and left (OS) eyes, with slight variations when compared to previous or subsequent clinic visits.
Conclusions:
- Drive-through iCare IOP measurements may overestimate IOP, necessitating confirmation with in-clinic measurements for significant changes.
- As teleophthalmology expands, optimizing the accuracy of drive-through tonometry is crucial for effective glaucoma care.
- Despite potential inaccuracies, drive-through tonometry remains a promising method for high-throughput, low-contact IOP monitoring.