The performance of saliva test strips for determining ethanol levels, as compared to gas chromatography and
Niklas Wargh1, Juha Piltti1, Pirjo Hedberg1
1NordLab Wellbeing Services, Oulu University Hospital, Oulu, Finland.
Abstract:
We set out to determine the performance of the Testi Technologies enzymatic assay saliva ethanol test strips of three different detection levels: 0 g/L, 0.2 g/L and 0.5 g/L, using as the reference method a gas chromatography analyser (GC). Alcohol levels were measured in 104 volunteers at up to three points in time, using up to three test strips per measurement, while gathering blood samples and breathalyser readings in parallel. The plasma alcohol concentrations (PAC) were determined from the plasma samples using GC. The qualitative results of the test strips were compared to the quantitative results from the reference method, as well as the breathalyser readings, and the amount of true and false positive and true and false negative results were classified using predetermined cut-off levels. The best performing test strips were the 0 g/L and the 0.2 g/L strips. The 0 g/L strips had a sensitivity and specificity of 1.00, as false negatives and false positives were not detected. The 0.2 g/L strips had a sensitivity and specificity [95% confidence interval (CI)] of 0.98 (0.96 - 1.00) and 0.83 (0.62 - 1.00) respectively, an accuracy of 0.97 (0.95 - 0.99), and a diagnostic odds ratio of 205.00 (35.33 - 1189.66). The test strips perform their intended purpose of screening for alcohol consumption well, with their great sensitivity as a defining property compared to other testing methods. For them to be able to be implemented in a clinical setting however, further refinement of the tests' characteristics would be required.
Related Concept Videos
Serum Laboratory Studies, Stool Test, Breath Test
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...