Related Experiment Video
Updated: Jul 16, 2025

Gradient Echo Quantum Memory in Warm Atomic Vapor
Published on: November 11, 2013
A simple method for rapid removal of the memory effect in cavity ring-down spectroscopy water isotope measurements
Jonathan Keinan1,2, Yonaton Goldsmith1
1Institute of Earth Sciences, The Hebrew University of Jerusalem, Jerusalem, Israel.
Rationale:
The accuracy determined in the routine analysis of water isotopes (δ17 O, δ18 O, δ2 H) using cavity ring-down spectroscopy is greatly affected by the memory effect (ME), a sample-to-sample carryover that biases measurements. This study aims to develop a simple method that rapidly removes the ME.
Methods:
We developed a method, designed for the Picarro L2140-i, that removes the ME by injecting small amounts of water with an extreme isotopic value ("kick") in the opposite direction of the ME. We conducted 11 experiments to identify the optimal kick for pairs of isotopically enriched and depleted samples. Once quantified, the optimal kick was used to create an ME-free, unbiased calibration curve, which was verified using international and internal lab standards.
Results:
Our kick method removes the ME very efficiently in half the time it takes for experiments without a kick. The optimal number of kick injections required to minimize stabilization time between standards of different compositions is three injections of δ2 H ≈ -1000‰ water per a 100‰ difference between standards. Three runs of routine measurements using the kick method resulted in uncertainties of 0.03‰, 0.2‰, and 5 permeg for δ18 O, δ2 H, and 17 O-excess, respectively.
Conclusions:
This study demonstrates a new method for rapidly removing the ME. Our kick protocol is a readily available, cheap, and efficient approach to reduce instrumental bias and improve measurement accuracy.
Related Concept Videos
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
¹H NMR of Labile Protons: Deuterium (²H) Substitution
Mass Spectrometry: Isotope Effect