Mutual interference between 3,6-dichlorinated carbazole and p,p'-dichlorodiphenyltrichloroethane in gas
Shanshan Zhou1, Qi Su2, Dan Zhong2
1College of Environment, Zhejiang University of Technology, Hangzhou, 310014, China. sszhou@zjut.edu.cn.
Abstract:
The widespread contamination of the environment by polyhalogenated carbazoles (PHCZs) has been increasingly observed during the past decade. Among numerous PHCZ congeners, 3,6-dichlorocarbazole (36-CCZ) is often among the most frequently detected at higher concentrations. Although the environmental level of the legacy pesticide p,p'-dichlorodiphenyltrichloroethane (p,p'-DDT) has been declining, it continues to be ubiquitously detected. These two compounds were found to interfere with each other during analyses using gas chromatography (GC) coupled with single- or triple-quadrupole low-resolution mass spectrometry (MS or MS/MS). The base peak in the mass spectra was that of m/z 235 for both compounds. In MS/MS with multiple reaction monitoring (MRM), the same transitions (235 → 200 and 235 → 165) were often used. Under the same GC operating conditions, the SH-I-5MS capillary column used in this work did not resolve the two compounds at baseline. Pre-treatment using cleanup column chromatography can fractionate the sample extract, with the two compounds separated in different fractions before instrumental analyses. Reversed-phase HPLC columns also work for resolving 36-CCZ and p,p'-DDT. Possible overlaps in GC retention and similarity in MS spectra might have caused data inaccuracy for 36-CCZ as well as p,p'-DDT in some studies published to date, and steps to avoid the interference should be taken into quality control protocols in future research and environmental monitoring.
Related Concept Videos
Mass Spectrometry: Cycloalkane Fragmentation
For example, cyclohexane molecular ions have a mass-to-charge ratio (m/z) of 84, which tends to produce a stronger signal than linear alkanes like hexane. This stability comes from...
Chemical Ionization (CI) Mass Spectrometry
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Mass Spectrometry: Alkyl Halide Fragmentation
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Types of Detectors-II