Gaussian optics optimization of an optical particle counter for measurement in high-pressure gas
Abstract:
An optimization model for the parameters of the pressure-resistant lens of an optical particle counter, based on Gaussian optics theory, was established to increase the measurement accuracy of the counter for high-pressure natural gas. Comparing the experimental and calculated values of the calibrated model, when the pressure-resistant lens is displaced by 2 mm under atmospheric pressure, the relative error of the measured body deformation is 0.15%. When the air pressure varies in the range 0.10-5.09 MPa, the maximum relative error of optical measurement volume deformation with the change of refractive index is 0.13%, which shows that the model has high reliability and accuracy.