Practical study on the impact of injection conditions in gradient elution mode for the analysis of therapeutic
Raquel Pérez-Robles1, Szabolcs Fekete2, Natalia Navas3
1Instituto de Investigación Biosanitaria de Granada (ibs.GRANADA), Granada, Spain; Department of Analytical Chemistry, Science Faculty, University of Granada, Granada, Spain; Fundación para la Investigación Biosanitaria de Andalucía Oriental-Alejandro Otero, Granada, Spain.
Abstract:
The impact of injected sample volume on apparent efficiency has been studied for very short columns in a systematic way. For large molecules such as therapeutic proteins, it was found that relatively large volumes can be injected onto ultra-short RPLC and IEX columns (i.e. L < 50 mm) without significantly affecting the quality of the separation. This favourable behavior is due to the on-off elution mechanism of large molecules and to the fact that therapeutic protein samples are formulated in aqueous-based media, which is the weakest solvent in RPLC and IEX. Therefore, their peak is strongly focused at the column inlet even when large volume is injected, and pre-column peak dispersion is compensated. However, ultra-short HILIC columns do not seem to be favorable, as they require for very low injection volume to avoid detrimental peak splitting and breakthrough effects. Such peak distortion is related to the inherent solvent mismatch between sample diluent (aqueous) and mobile phase strength (highly organic in HILIC). When studying mass load, the ranking of the elution modes was the same, and the largest relative mass could be injected onto IEX columns (as large as 10% sample to sorbent mass), without affecting the separation quality.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
10:21Automated Hydrophobic Interaction Chromatography Column Selection for Use in Protein Purification
Published on: September 21, 2011
Related Concept Videos
High-Performance Liquid Chromatography: Elution Process
Principles Of Column Chromatography
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Types Of Column Chromatography
Gel Filtration Chromatography
When the...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...