Multi-wavelength deep-ultraviolet absorbance detector based upon program-controlled pulsing light-emitting diodes
Ibraam E Mikhail1, Mohamed Hemida1, Leo Lebanov1
1Australian Centre for Research on Separation Science (ACROSS), School of Natural Sciences (Chemistry), University of Tasmania, Hobart, Tasmania 7001, Australia.
A new pulsed multi-wavelength ultraviolet absorbance detector (PMLAD) was developed using a single board computer and FPGA. This novel system enables simultaneous analysis of pharmaceuticals like paracetamol, caffeine, and aspirin with high accuracy.
Area of Science:
- Analytical Chemistry
- Spectroscopy
- Instrumentation
Background:
- Traditional UV absorbance detection methods can be limited in speed and multiplexing capabilities.
- Deep-ultraviolet light-emitting diodes (DUV-LEDs) offer potential for compact and efficient absorbance measurements.
- Integrating advanced computing with optical detection is crucial for modern analytical instrumentation.
Purpose of the Study:
- To introduce a novel pulsed multi-wavelength ultraviolet absorbance detector (PMLAD) for enhanced analytical performance.
- To demonstrate the system's capability for simultaneous pharmaceutical analysis using liquid chromatography.
- To validate the detector's linearity, accuracy, and quantitation limits.
Main Methods:
- A single board computer (SBC) with a field-programmable gate array (FPGA) was used to generate synchronized micro-pulses for three DUV-LEDs (235, 250, 280 nm).
- Data acquisition and processing were performed using a custom-made program, with combined DUV-LED outputs transmitted via optical fiber to a flow cell.
- An ultra-fast photodiode recorded DUV pulse intensities, and the system was coupled with a standard liquid chromatograph.
Main Results:
- The PMLAD demonstrated excellent linearity with upper limits of linearity (A95%) ranging from 1768 to 2189 mAU and effective path lengths ≥98.0%.
- Stray light was measured to be ≤0.9%.
- Simultaneous determination of paracetamol, caffeine, and aspirin was achieved with high linearity (r² ≥ 0.99996) and low quantitation limits (LOQs) down to 0.10 µg/mL.
Conclusions:
- The developed pulsed multi-LEDs absorbance detector (PMLAD) offers a sensitive and selective method for pharmaceutical analysis.
- The use of absorbance ratios between wavelengths aids in compound identification.
- This novel approach provides a robust platform for multi-wavelength UV detection in chromatography.