Physical characteristics of deep learning-based image processing software in computed tomography: a phantom study
Seiya Sato1, Atsushi Urikura2, Makoto Mimatsu1
1Department of Radiological Technology, Radiological Diagnosis, National Cancer Center Hospital, 5-1-1 Tsukiji, Chuo-Ku, Tokyo, 104-0045, Japan.
Purpose:
This study aimed to assess the image characteristics of deep-learning-based image processing software (DLIP; FCT PixelShine, FUJIFILM, Tokyo, Japan) and compare it with filtered back projection (FBP), model-based iterative reconstruction (MBIR), and deep-learning-based reconstruction (DLR).
Methods:
This phantom study assessed the object-specific spatial resolution (task-based transfer function [TTF]), noise characteristics (noise power spectrum [NPS]), and low-contrast detectability (low-contrast object-specific contrast-to-noise ratio [CNRLO]) at three different output doses (standard: 10 mGy; low: 3.9 mGy; ultralow: 2.0 mGy). The processing strength of DLIPFBP with A1, A4, and A9 was compared with those of FBP, MBIR, and DLR.
Result:
The standard dose with high-contrast TTFs of DLIPFBP exceeded that of FBP. Low-contrast TTFs were comparable to or lower than that of FBP. The NPS peak frequency (fP) of DLIPFBP shifts to low spatial frequencies of up to 8.6% at ultralow doses compared to the standard FBP dose. MBIR shifted the most fP compared to FBP-a marked shift of up to 49%. DLIPFBP showed a CNRLO equal to or greater than that of DLR in standard or low doses. In contrast, the CNRLO of the DLIPFBP was equal to or lower than that of the DLR in ultralow doses.
Conclusion:
DLIPFBP reduced image noise while maintaining a resolution similar to commercially available MBIR and DLR. The slight spatial frequency shift of fP in DLIPFBP contributed to the noise texture degradation suppression. The NPS suppression in the low spatial frequency range effectively improved the low-contrast detectability.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET