Related Experiment Video
Updated: Jul 16, 2025

06:24
A Spine Robotic-Assisted Navigation System for Pedicle Screw Placement
Published on: May 11, 2020
8.9K
The pedicle screw accuracy using a robotic system and measured by a novel three-dimensional method
Marcelo Oppermann1,2, Vahagan Karapetyan3, Shaurya Gupta4
1Department of Clinical Neurological Science, Schulich School of Medicine and Dentistry, Western University, London, ON, Canada. Marcelo.Oppermann@lhsc.on.ca.
Journal of Orthopaedic Surgery and Research
|September 20, 2023
Summary
Robotic spine surgery accuracy was evaluated using the Mazor system. CT-to-fluoroscopy registration improved pedicle screw (PS) accuracy compared to preopCT/C-arm, though nearly half of PS insertions were non-optimal.
Area of Science:
- Spine surgery
- Medical robotics
- Surgical accuracy
Background:
- Robotic systems enhance precision in medicine, particularly in spine surgery.
- Shared-control robotic systems offer "steady-hand" manipulation.
- Assessing the true accuracy of robotic instrumentation is crucial.
Purpose of the Study:
- To evaluate the spatial accuracy of pedicle screw (PS) insertion using the Mazor robotic system.
- To compare the accuracy of different registration methods (preopCT/C-arm vs. CT-to-fluoroscopy).
- To analyze screw trajectory deviations and classify error patterns.
Main Methods:
- Two cadaveric specimens (C7-S1) were used for pedicle screw (PS) insertion.
- Planning involved two registration methods: preopCT/C-arm and CT-to-fluoroscopy.
- PS insertion was performed via open and percutaneous approaches, followed by CT acquisition for accuracy assessment.
Main Results:
- CT-to-fluoroscopy registration showed superior accuracy (0.9 mm deviation) compared to preopCT/C-arm (1.7 mm deviation, p < 0.003).
- Percutaneous access was slightly more accurate than open access (1.3 mm vs. 1.7 mm).
- While 10.8% of screws were unacceptable by quantitative 3D position, nearly half (49.3%) were considered non-optimal under a rigorous inaccuracy concept.
Conclusions:
- CT-to-fluoroscopy registration significantly enhances pedicle screw accuracy in robotic spine surgery.
- A substantial proportion of pedicle screw insertions may be non-optimal, necessitating careful accuracy assessment.
- The O-arm registration method proved more accurate than preopCT/C-arm for implant placement.