Related Experiment Video
Updated: Jul 16, 2025

Terahertz Microfluidic Sensing Using a Parallel-plate Waveguide Sensor
Published on: August 30, 2012
Real-time hybrid angular-interrogation surface plasmon resonance sensor in the near-infrared region for wide dynamic
Hidenori Koresawa1, Kota Seki2, Kenji Nishimoto2
1Graduate School of Advanced Technology and Science, Tokushima University, 2-1 Minami-Josanjima, Tokushima, Tokushima, 770-8506, Japan.
Abstract:
Herein, we integrated angle-scanning surface plasmon resonance (SPR) and angle-fixed SPR as a hybrid angular-interrogation SPR to enhance the sensing performance. Galvanometer-mirror-based beam angle scanning achieves a 100-Hz acquisition rate of both the angular SPR reflectance spectrum and the angle-fixed SPR reflectance, whereas the use of near-infrared light enhances the refractive index (RI) sensitivity, range, and precision compared with visible light. Simultaneous measurement of the angular SPR reflectance spectrum and angle-fixed SPR reflectance boosts the RI change range, RI resolution, and RI accuracy to 10-1-10-6 RIU, 2.24 × 10-6 RIU, and 5.22 × 10-6 RIU, respectively. The proposed hybrid SPR is a powerful tool for wide-dynamic-range RI sensing with various applications.