Procedure and Key Optimization Strategies for an Automated Capillary Electrophoretic-based Immunoassay Method
Published on: September 10, 2017
The application of capability indices in the validation of ELISA methodology
1Piramal Pharma Solutions, Analytical Development, Earls Road, Grangemouth, Stirlingshire, FK3 8XG, UK.
Abstract:
Aim: To investigate the impact of sample replication number (duplicate vs triplicate) on the validation of ELISA methodology. Materials & methods: The methodology was validated with reference sample and test sample as an 11-point triplicate dilution series. The data were reanalyzed post-validation as if conducted as a duplicate dilution series. Results: The triplicate methodology was validated with a precision of 5.3% and mean bias of -1.7%. The duplicate methodology generated a precision of 5.7% and mean bias of -2.2%. Conclusion: Both the triplicate (method capability index = 1.37) and duplicate (method capability index = 1.25) ELISA methodology can support an 80-125% relative potency specification with a 0.004% or 0.018% probability of out-of-specification results, respectively.
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...
Data Validation
Key parameters for method validation include: