Point-of-Care Ultrasound: A Review of Ultrasound Parameters for Predicting Difficult Airways
Published on: April 7, 2023
Role of ultrasonography in upper airway assessment: A comparison with computerized tomography measurements
Muhammedcan Sen1, Muge Gunalp1, Ahmet Burak Oguz1
1Ankara University School of Medicine, Department of Emergency Medicine, Ankara, Turkey.
Introduction:
This study aimed to compare ultrasonography (US) measurements of the upper airway to computerized tomography (CT) measurements. Our study's primary outcome is to research the accuracy of US measurements in the evaluation of upper airway diameters when CT is taken as the gold standard; the secondary outcome is to determine the time required to obtain US measurements.
Methods:
This prospective study included patients ≥18 years old that had undergone thoracic or neck CT due to current clinical necessity. The US measurement for each patient was performed by two researchers with different levels of experience, both of whom were blinded to each other and the CT measurements. Measurements were obtained from the vocal cords and subglottic region. The duration of the US performance was also recorded.
Results:
The US and CT measurements were obtained from 94 patients. Concordance between US and CT measurements was found in the Bland-Altman analysis, with a mean -0.05 mm difference for vocal cord diameter and -1.2 mm for subglottic diameter. The intra-class correlation coefficients (ICC) between the CT and US measurements were 0.993, and 0.609 for vocal cord and subglottic diameter measurements, respectively. The ICC between US performers was 0.992 for vocal cord diameter and 0.959 for subglottic diameter. The US's mean time for vocal cord diameter measurement was 38 ± 23 s, and the mean time for subglottic diameter measurement was 49 ± 30 s.
Conclusion:
The concordance between US and CT measurements is high and the measurements of different practitioners with different experience levels are compatible with each other.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies II: Ultrasonography
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...