Related Experiment Video
Updated: Jul 16, 2025

Dual-Dye Optical Mapping of Hearts from RyR2R2474S Knock-In Mice of Catecholaminergic Polymorphic Ventricular Tachycardia
Published on: December 22, 2023
ICD outcome in pediatric arrhythmogenic cardiomyopathy
Marianna Cicenia1, Massimo Stefano Silvetti1, Nicoletta Cantarutti1
1Pediatric Cardiology and Arrhythmia/Syncope Complex Unit, Bambino Gesù Children's Hospital, IRCCS, Rome, Italy.
Background:
Arrhythmogenic cardiomyopathy (ACM) is a very rare condition among pediatric patients. Sudden cardiac death (SCD) is the main complication and often requires ICD implantation. Aim of the study is the evaluation of the outcomes of ICD implanted ACM pediatric patients in terms of safety, efficacy and complications.
Methods:
All pediatric patients (<18 y.o.) diagnosed with ACM and who were implanted with ICD since 2009 in Our Institution were collected. Implantation was decided according to current recommendations/ guidelines, and outcome was recorded during follow-up.
Results:
Nineteen consecutive ACM patients were implanted with ICD. Subcutaneous ICDs (S-ICD) were implanted in 15 patients (79%) while transvenous ICDs (TV-ICD) in 4 patients (21%). Mean age at implantation was 14.3 ± 2.1 y.o. ICDs were implanted for secondary prevention in 4 (21%) patients, and for primary prevention in 15 (79%). During the follow-up (5.59 ± 3.4 years), appropriate ICD interventions were delivered in 4 (21%) patients for sustained VTs, [2 implanted in primary prevention (13%) and 2 in secondary prevention (50%)]. No defibrillation failures occurred. Inappropriate shocks occurred in 2 cases (10.5%). Device-related complications requiring device revision occurred in 3 (16%): lead dislodgement, surgical skin erosion and sensing defect.
Conclusions:
In a pediatric ACM cohort, appropriate ICD therapies occurred in a minority of primary prevention patients and frequently in secondary prevention patients. The rate of inappropriate shocks and device-related complications were even more rare and mostly wound related. Therefore, ICD therapy in pediatric ACM is effective and safe.
Related Concept Videos
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy V: Interprofessional Care
Cardiomyopathy IV: Restrictive Cardiomyopathy
Cardiomyopathy I: Introduction and Classification
Cardiomyopathy III: Hypertrophic Cardiomyopathy
Mechanism of Cardiac Arrhythmias