Related Experiment Video
Updated: Jul 15, 2025

Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
Safety of an esophageal deviator for atrial fibrillation catheter ablation
Renner Pereira1, Cristiano Pisani1, Vera Aiello1
1Arrhythmia Unit, Heart Institute, University of São Paulo Medical School, São Paulo, Brazil.
Background:
Esophageal thermal injury is a complication of atrial fibrillation (AF) ablation, and it can be avoided by esophageal deviation during left atrial posterior wall radiofrequency catheter ablation.
Objective:
This study aimed to evaluate the safety of a nitinol-based mechanical esophageal displacement device (MEDD) and its performance.
Methods:
This preclinical safety study was conducted on 20 pigs, with 10 undergoing radiofrequency AF ablation using the MEDD and 10 serving as a control group under anticoagulation but without radiofrequency application. Esophageal traumatic injuries were classified from 0 to 4 and were grouped as absent (grade 0), minor (grade 1 or 2), moderate (grade 3), or major risk lesions (grade 4) by anatomopathological study. Grades 1 and 2 were considered acceptable. Fluoroscopy was used to measure displacement.
Results:
Five (25%) pigs developed traumatic lesions, 4 with grade 1 and 1 with grade 2 (2-mm superficial ulcer). There was no difference in lesion occurrence between the radiofrequency and control groups (30% and 20%, respectively; P = .43). Under rightward displacement, the right edge moved 23.9 (interquartile range [IQR] 21.3-26.3) mm and the left edge moved 16.3 (IQR 13.8-18.4) mm (P < .001) from baseline. Under leftward displacement, the right edge moved 13.5 (IQR 10.9-15.3) mm and the left edge moved 16.5 (IQR 12.3-18.5) mm (P = .07). A perforation to the pharyngeal diverticulum occurred in 1 pig, related to an accidental extubation.
Conclusion:
In pigs, the MEDD demonstrated safety in relation to esophageal tissue, and successful deviation. Esophageal traumatic injuries were acceptable, but improper manipulation led to pharyngeal lesion.
More Related Videos
04:58Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
05:43Author Spotlight: Improving Lesion Contiguity in Pulmonary Vein Isolation via Proactive Esophageal Cooling
Published on: April 19, 2024
Related Concept Videos
Hemodialysis I: Introduction
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization IV: Nursing Management
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias