Related Experiment Video
Updated: Jul 15, 2025

Evaluating Regional Pulmonary Deposition using Patient-Specific 3D Printed Lung Models
Published on: November 11, 2020
[Research and Design of Calibration Device for Medical Test Lung]
Tingting Zhang1, Qiu Huang2, Xuhui Liao2
1Qinghai Red Cross Hospital, Xining, 810000.
Abstract:
In order to solve the problems of quality control and traceability of medical test lung for meeting the calibration conditions of JJF 1234-2018 Calibration Specification for Ventilators, the calibration device and method are researched for compliance and airway resistance of medical test lung in this paper. A calibration device for medical test lung is designed using constant volume active piston technology to simulate human breathing. Through comparison experiment, the deviation between this device and the similar foreign device can be found. The deviation is lower than 0.4% for lung compliance and lower than 0.7% for airway resistance. The calibration of lung compliance and airway resistance can be completed by this device. This device has a clear and complete traceability path to ensure quality control from the source. The calibration of ventilator is improved. This paper provides a reference for related metrology departments and medical institutions to study on quality inspection of respiratory medical instruments.
More Related Videos
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Lung Capacity