Volume-targeted mask ventilation during simulated neonatal resuscitation
Brenda Hiu Yan Law1,2, Georg M Schmölzer3,2
1Department of Pediatrics, University of Alberta, Edmonton, Alberta, Canada.
Objective:
Mask positive pressure ventilation (PPV) in the delivery room is routinely delivered with set peak inflation pressures. To aid mask PPV, stand-alone respiratory function monitors (RFMs) have been used in the delivery room, while ventilator-based, volume-targeted ventilation (VTV) is routinely used in the neonatal intensive care unit (NICU).
Design:
This is a prospective, randomised, crossover simulation study. Participants were briefly trained to use a neonatal ventilator for volume-targeted mask ventilation (VTV-PPV), then performed mask ventilation on a manikin in a randomised order using VTV-PPV, T-piece PPV or T-piece PPV with RFM visible.
Setting:
In situ in a neonatal resuscitation room within a level 3 NICU.
Participants:
Healthcare professionals (HCPs) trained in neonatal resuscitation with experience as team leaders.
Interventions:
Semiautomated, ventilator-based VTV-PPV using two-hand hold versus manual PPV via a T-piece device (T-piece, RFM masked) versus manual PPV with RFM visible using one-hand hold.
Main Outcome Measures:
Respiratory characteristics including % mask leak, tidal volume (VT) and peak inflation pressure (PIP).
Results:
Thirty-two HCPs (23 (72%) female and 9 (28%) male) participated. The median mask leak was significantly lower with 'VTV-PPV' (11%, IQR 0%-14%) compared with both 'T-piece, RFM visible' (82%, IQR 30%-91%) and 'T-piece, RFM masked' (81%, IQR 47%-91%) (p<0.0001). The median delivered VT was 4.1 mL/kg (IQR 3.9-4.4) with VTV-PPV compared with 2.1 mL/kg (IQR 1.2-9) with T-piece, RFM visible and 1.8 mL/kg (IQR 1.1-5.8) with T-piece, RFM masked (p=0.0496). PIP was also significantly lower with VTV-PPV.
Conclusion:
During neonatal simulation, VTV-PPV reduced mask leak and allowed for consistent VT delivery compared with T-piece with and without RFM guidance.
More Related Videos
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...