Measurement accuracy of a microwave doppler sensor beneath the mattress as a continuous respiratory rate monitor: a
Hiroyuki Tanaka1, Masashi Yokose2, Shunsuke Takaki1
1Department of Anesthesiology and Critical Care Medicine, Yokohama City University Graduate School of Medicine, 3-9 Fukuura, Kanazawa-ku, Yokohama, Japan.
Purpose:
Non-contact continuous respiratory rate monitoring is preferred for early detection of patient deterioration. However, this technique is under development; a gold standard respiratory monitor has not been established. Therefore, this prospective observational method comparison study aimed to compare the measurement accuracy of a non-contact continuous respiratory rate monitor, a microwave Doppler sensor positioned beneath the mattress, with that of other monitors.
Methods:
The respiratory rate of intensive care unit patients was simultaneously measured using a microwave Doppler sensor, capnography, thoracic impedance pneumography, and a piezoelectric sensor beneath the mattress. Bias and 95% limits of agreement between the respiratory rate measured using capnography (standard reference) and that measured using the other three methods were calculated using Bland-Altman analysis for repeated measures. Clarke error grid (CEG) analysis evaluated the sensor's ability to assist in correct clinical decision-making.
Results:
Eighteen participants were included, and 2,307 data points were analyzed. The bias values (95% limits of agreement) of the microwave Doppler sensor, thoracic impedance pneumography, and piezoelectric sensor were 0.2 (- 4.8 to 5.2), 1.5 (- 4.4 to 7.4), and 0.4 (- 4.0 to 4.8) breaths per minute, respectively. Clinical decisions evaluated using CEG analyses were correct 98.1% of the time for the microwave Doppler sensor, which was similar to the performance of the other devices.
Conclusion:
The microwave Doppler sensor had a small bias but relatively low precision, similar to other devices. In CEG analyses, the risk of each monitor leading to inadequate clinical decision-making was low.
Trial Registration Number:
UMIN000038900, February 1, 2020.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...