Assessment of the effectiveness of Saba shielding with the composition of Cu-Bi in neck CT imaging: a phantom and
Mohammad Keshtkar1, Mohammad Amin Mosleh-Shirazi2,3, Valiallah Saba4
1Medical Physics and Radiology Department, Faculty of Medicine, Gonabad University of Medical Sciences, Gonabad, Iran.
Abstract:
The use of computed tomography (CT) is a very well-established medical diagnostic imaging modality, however, the high radiation dose due to this imaging method is a major concern. Therefore, dose reduction methods are necessary, especially for superficial radiosensitive organs like the thyroid. The aim of this study is to construct and assess a CT shield with composition of 90% Cu and 10% Bi (Saba shield) with regard to dose reduction and image quality. The efficiency of the constructed shields for dose reduction was assessed by measuring entrance skin dose (ESD), using thermoluminescence dosimeters placed on an anthropomorphic phantom. Image quality was assessed quantitatively based on image noise and CT number accuracy by drawing regions of interest on CT images of the anthropomorphic phantom. Image quality was further investigated qualitatively in a patient study. Application of the Saba shield and 100% Bi shield with the thickness of one thickness (1T) reduced ESD by 50.2% and 51.7%, respectively, and using a three-fold thickness reduced ESD by 64.6% and 65.1%, respectively. Saba shield with thickness of 1T had no significant change in image noise in the anterior part, and image noise and mean CT number in the posterior part (P> 0.05). The statistical analysis performed did not find any meaningful difference between the study and control groups in image quality assessment of the patient study (P> 0.05). The 1T Saba shield reduced thyroid dose efficiently during neck CT imaging without causing unwanted effects on image quality.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography