Capillary electrophoresis in parenteral nutrition control - validation of two analytical methods: Amino
Camille Fanjeaux1, Raphael Vazquez2, Marie-Noëlle Guerrault-Moro1
1pharmacie, CHI Poissy-Saint-Germain-en-Laye, Poissy, France.
Objectives:
Production of parenteral nutrition bags (PNBs) involves many nutrients: complete control of the production process decreases the risk of error. This study aimed to develop and validate two analytical methods by capillary electrophoresis (CE) for simultaneous detection of: glucose, amino acids (Primene®) and glucose-1-phosphate (Phocytan®) (anionic method, AM) on one hand; and on the other hand potassium, sodium, calcium and magnesium (cationic method, CM).
Methods:
Methods were developed using capillary electrophoresis with diode array detection (CE-DAD) (CE 7100, Agilent), indirect photometric detection, 56 cm long capillary and two different buffers (pH=12.1 for AM and pH=3.2 for CM). These methods were validated according to guidelines from the Société Française des Sciences et Techniques Pharmaceutiques (SFSTP).Analytical parameters were optimised: temperature was regulated at 15°C and the current settled to - 15kV, for a 21 minute analysis time for AM. Conditions were settled to 25°C and 30kV for CM so the analysis time dropped to 7 minutes.Accuracy profiles were established and recovery rates (RR), Repeatability and Reproducibility Coefficient of Variation (respectively RaCV and RoCV) were calculated.Capability was also calculated for each nutrient and concentration range according to guidelines from the Evaluation and Research Group on Protection in a Controlled Atmosphere (GERPAC).
Results:
Methods were successfully validated with: RR between 99.2 and 101.9%, RaCV between 1.5 and 3.1%, and RoCV between 2.4 and 4.1% for AM, and RR between 97.5 and 102.7%, RaCV between 0.5 and 2.3%, and RoCV between 0.6 and 2.8% for CM.Accuracy profiles were established with 95% β probability, except for glucose-1-phosphate (90%). Acceptance limits were settled to ±1 0% of target value. Capabilities are defined as "good" or "very good".
Conclusions:
The methods developed by this research will ensure the composition of PNB is compliant to PNB formulas. These results show CE is an appropriate method for PNB quantitative control.CE utilisation for controlling other hospital preparations seems to be a relevant alternative to conventional methods such as liquid chromatography.
More Related Videos
00:04A Clinical Trial Assessing the Safety, Efficacy, and Delivery of Olive-Oil-Based Three-Chamber Bags for Parenteral Nutrition
Published on: September 20, 2019
07:46Sheathless Capillary Electrophoresis–Mass Spectrometry for Metabolic Profiling of Biological Samples
Published on: October 1, 2016
Related Concept Videos
Parentral Nutrition: Centeral and Peripheral Parental Nutrition
PN can be administered through two primary routes:
1. Central Parenteral Nutrition (CPN):
CPN involves delivering a high concentration of nutrients through a large vein. This is typically achieved using a Peripherally Inserted Central Catheter (PICC) or,...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Capillary Electrophoresis: Instrumentation