Speech production real-time MRI at 0.55 T
Yongwan Lim1, Prakash Kumar1, Krishna S Nayak1
1Ming Hsieh Department of Electrical and Computer Engineering, Viterbi School of Engineering, University of Southern California, Los Angeles, California, USA.
Purpose:
To demonstrate speech-production real-time MRI (RT-MRI) using a contemporary 0.55T system, and to identify opportunities for improved performance compared with conventional field strengths.
Methods:
Experiments were performed on healthy adult volunteers using a 0.55T MRI system with high-performance gradients and a custom 8-channel upper airway coil. Imaging was performed using spiral-based balanced SSFP and gradient-recalled echo (GRE) pulse sequences using a temporal finite-difference constrained reconstruction. Speech-production RT-MRI was performed with three spiral readout durations (8.90, 5.58, and 3.48 ms) to determine trade-offs with respect to articulator contrast, blurring, banding artifacts, and overall image quality.
Results:
Both spiral GRE and bSSFP captured tongue boundary dynamics during rapid consonant-vowel syllables. Although bSSFP provided substantially higher SNR in all vocal tract articulators than GRE, it suffered from banding artifacts at TR > 10.9 ms. Spiral bSSFP with the shortest readout duration (3.48 ms, TR = 5.30 ms) had the best image quality, with a 1.54-times boost in SNR compared with an equivalent GRE sequence. Longer readout durations led to increased SNR efficiency and blurring in both bSSFP and GRE.
Conclusion:
High-performance 0.55T MRI systems can be used for speech-production RT-MRI. Spiral bSSFP can be used without suffering from banding artifacts in vocal tract articulators, provide better SNR efficiency, and have better image quality than what is typically achieved at 1.5 T or 3 T.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...