Related Experiment Video
Updated: Jul 28, 2026

Four-Dimensional CT Analysis Using Sequential 3D-3D Registration
Published on: November 23, 2019
[A 3D/2D registration method based on reconstruction of orthogonal-view Xray images]
1School of Biomedical Engineering, Southern Medical University//Guangdong Provincial Key Laboratory of Medical Image Processing//Guangdong Province Engineering Laboratory for Medical Imaging and Diagnostic Technology, Southern Medical University, Guangzhou 510515, China.
Objective:
To establish a 3D/2D registration method for preoperative CT and intra-operative X-ray images in imageguided spine surgery.
Methods:
We propose a 3D/2D registration algorithm based on 3D image reconstruction. The algorithm performs 3D image reconstruction of 2D orthogonal view X-ray images, thus converting the problem into 3D/3D registration. By constructing an end-to-end framework that combines the two tasks of reconstruction and registration, the geodesic distance is measured in the 3D manifold space to complete the registration.
Results:
We conducted experiments on the public dataset CTSpine1k. The tests on two test sets with different initial registration errors showed that for data with small initial errors, the proposed algorithm achieved a rotation estimation error of 0.115±0.095° and a translation estimation error of 0.144±0.124 mm; for data with larger initial errors, a rotation estimation error of 0.792±0.659° and a translation estimation error of 0.867±0.701 mm were achieved.
Conclusion:
The proposed method can achieve robust and accurate 3D/2D registration at a speed that meets real-time requirements to improve the performance of spine surgery navigation.