EURADOS intercomparison IC2020ph on whole body dosemeters for photons
Argiro Boziari1, Hannes Stadtmann2, Panagiotis Askounis1
1Dosimetry and Calibration Unit, Greek Atomic Energy Commission, Agia Paraskevi 15341, Greece.
Abstract:
The European Dosimetry Group has been organising international intercomparisons (ICs) for personal dosemeters on a regular basis. The IC announced in 2020 was postponed to the end of 2020 and the beginning of 2021 due to coronavirus (COVID-19) restrictions. The irradiation plan consisted of nine irradiation setups with five different photon radiation qualities (S-Cs, S-Co, N-150, W-60, W-80) and two different angles of irradiation incidence (0o and 60o). A total of 112 monitoring services from 50 countries with 132 dosimetry systems participated. The present work describes and analyses the individual results for the personal dose equivalent quantities Hp(10) and, if submitted, Hp(0.07) for all participating systems. It also presents the organisational details and the difficulties arising from the COVID-19 crisis during the IC. The statistical results showed a satisfactory performance with the medians of all Hp(10) and Hp(0.07) response values very close to unity. However, there are few individual monitoring services with significant large variations.