Purpose:
To investigate the resistance of an implant crown telescopically engaged to a geometrically defined hexagonal abutment with and without cement under compressive dynamic cyclic fatigue loading.
Materials And Methods:
40 crowns, 9 mm in height, were cast from prefabricated plastic copings, each telescopically engaged to a 3-mm-high machined hexagonal abutment. 20 crowns were retained with zinc phosphate cement, and 20 crowns were retained with a screw. A vertical load of 15 kg was applied to all samples under oscilla)on. A Periotest was used to measure the mobility of the implant crown in Periotest values (PTV). Test endpoints were defined by the following: fatigue cycles ≥ 20 million; crown PTV > 10; or if samples became visibly loose or component fracture.
Results:
Cement-retained crowns failed on average at 2.60 x 106 cycles ± 2.27, while screw-retained crown samples failed at 2.17 x 106 cycles ± 1.27 with no significant difference (P > .05). Implant and abutment screw fractures were the most prevalent mode of failure in the cement-retained group, while in the screw-retained group, failures were caused by the loosening of one or both screw joints. The rate of increase in PTV was higher in the screw-retained group than in the cement-retained group.
Conclusions:
Under the experimental conditions, an implant crown telescopically engaged vertically to a 3-mm-tall hexagonal abutment, under compressive dynamic cyclic fatigue loading with or without cement, demonstrated no differences in resistance or failure outcomes. A cement-retained implant crown with telescopic engagement to the abutment is more rigid, resulting in more implant and abutment screw fractures than loosening.
Related Concept Videos
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in...
Behavior of Concrete Under Compressive Load
As the concrete specimen fractures under...
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Normal Strain under Axial Loading
Fatigue Strength of Concrete
Fatigue