Optical Quality and Higher Order Aberrations of Refractive Extended Depth of Focus Intraocular Lenses
Purpose:
To compare the optical quality and higher order aberrations of four different refractive extended depth of focus intraocular lens (EDOF IOL) models (Lentis Comfort [Teleon Surgical BV], MiniWell (SIFI), LuxSmart [Bausch & Lomb], and AcrySof IQ Vivity [Alcon Laboratories, Inc]) and a monofocal IOL (Tecnis ZCB00; Johnson & Johnson Surgical Vision, Inc).
Methods:
An optical metrology instrument (OptiSpheric IOL PRO2; Trioptics GmbH) was used to study image quality metrics of the different IOLs. The modulation transfer function (MTF) was measured at pupil sizes 1 to 5 mm in 1-mm steps. Area under the MTF and simulated visual acuity were calculated for different pupil diameters. Wavefront aberrations were measured with the SHS Ophthalmic device (Optocraft GmbH), which features a Hartmann-Shack sensor.
Results:
All EDOF lenses yielded a simulated far visual acuity of 0.00 logMAR (20/20 Snellen) or better. At the 0.20 logMAR visual acuity level, the EDOF IOLs showed an increased depth of focus of at least 0.75 diopters compared to the monofocal IOL. Pupil dependency was more pronounced with the MiniWell, LuxSmart, and Vivity, whereas the Lentis showed a more consistent behavior at different apertures. The wavefront measurement revealed increased central aberrations for the MiniWell, LuxSmart, and Vivity IOL compared to the monofocal control.
Conclusions:
All EDOF IOLs demonstrated increased depth of focus while maintaining good simulated visual acuity at the far focus. Pupil dependency differed between the IOL models. The wavefront analysis revealed the complex design of the different EDOF IOLs with several zones with varying aberrations. [.
More Related Videos
Related Concept Videos
Focusing of Light in the Eye
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Glaucoma: Overview
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment