Performances and limits of Bag-Valve-Device for pre-oxygenation and manual ventilation: A comparative bench and
1Ventilation Laboratory (Vent'Lab), Medical Intensive Care Unit (ICU), Angers University Hospital, Angers, France; Med(2)Lab, Air Liquide Medical Systems, Antony, France.
Introduction:
Bag-Valve-Device (BVD) is the most frequently used device for pre-oxygenation and ventilation during cardiopulmonary resuscitation (CPR). A minimal expired fraction of oxygen (FeO2) above 0.85 is recommended during pre-oxygenation while insufflated volume (VTi) should be reduced during manual ventilation. The objective was to compare the performances of different BVD in simulated conditions.
Methods:
Nine BVD were evaluated during pre-oxygenation: spontaneous breathing patients were simulated on a test lung (mild and severe conditions). FeO2 was measured with and without positive end-expiratory pressure (PEEP). CO2 rebreathing was evaluated. Then, manual ventilation was performed by 36 caregivers (n = 36) from three hospitals on a specific manikin; same procedure was repeated by 3 caregivers (n = 3) on two human cadavers with three of the nine BVD: In non-CPR scenario and during mechanical CPR with Interrupted Chest Compressions strategy (30:2).
Results:
Pre-oxygenation: FeO2 was lower than 0.85 for three BVD in severe condition and for two BVD in mild condition. FeO2 was higher than 0.85 in eight of nine BVD with an additional PEEP valve (PEEP 5 cmH2O). One BVD induced CO2 rebreathing. Manual ventilation: For non-CPR manual ventilation, mean VTi was within the predefined lung protective range (4-8 mL/kg PBW) for all BVD on the bench. For CPR manual ventilation, mean VTi was above the range for three BVD on the bench. Similar results were observed on cadavers.
Conclusions:
Several BVD did not reach the FeO2 required during pre-oxygenation. Manual ventilation was significantly less protective in three BVD. These observations are related to the different BVD working principles.
More Related Videos
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
09:14Establishment of a Novel Ex Vivo Lung Perfusion System for Rat Lungs After Circulatory Death
Published on: October 18, 2024
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: