Related Experiment Video
Updated: Jul 13, 2025

06:44
Cantilever Bending of Murine Femoral Necks
Published on: January 5, 2022
2.2K
Bending 30-gauge needles using a needle guide: fatigue life evaluation.
Jared Joseph Tuttle1, Andrew Doran Davidson2, Gregory Kent Tuttle3
1University of Texas Health Science Center at San Antonio, Joe R. and Teresa Lozano Long School of Medicine, San Antonio, TX, USA.
Journal of Dental Anesthesia and Pain Medicine
|October 16, 2023
Summary
Using a needle guide significantly increases the fatigue life of 30-gauge dental needles, allowing for an average of 40 bends before fracture. This enhances safety for dental professionals handling sharps and performing injections.
Area of Science:
- Dental materials science
- Biomedical engineering
- Medical device safety
Background:
- Dentists often bend needles for injections, raising concerns about fracture, lumen occlusion, and sharps handling.
- Previous research indicated 30-gauge needles fracture after 4-9 bends without a guide.
Purpose of the Study:
- To evaluate the fatigue life of 30-gauge dental needles when bent using a needle guide.
- To determine the number of 90° bends a needle can sustain before fracture.
Main Methods:
- 48 30-gauge needles were tested by two operators using a needle guide.
- Needles were bent to 90°, tested for anesthetic flow, then returned to 0° until failure.
- Failure was defined as anesthetic not passing through the needle tip due to fracture or occlusion.
Main Results:
- Needles sustained an average of 40.33 bends (minimum 20, median 40, maximum 54) before fracture.
- The needle lumen remained patent until fracture occurred in all trials.
- Operator differences were statistically significant (P < 0.001), but lot differences were not (P = 0.504).
Conclusions:
- Needle guides significantly increase the number of bends before needle fracture compared to previous studies (P < 0.000001).
- Further research should explore needle guides with other needle types and various operators.
- Investigating workplace safety and clinical applications of bent needles for anesthetic delivery is warranted.
Related Concept Videos
Fatigue
189
Fatigue occurs when materials rupture under repeated or fluctuating loads, even at stress levels far below their static breaking strength. It typically results in brittle failure, even for ductile materials. It is a critical consideration in designing machines and structural components subjected to repetitive or varying loads. The nature of these loadings can range from fluctuating loads like unbalanced pump impellers causing vibrations to repeatedly bending a thin steel rod wire back and forth...
189
Bending of Material: Problem Solving
191
In this lesson, determine the ratio of the maximum bending moments applied to two metal pipes, given that both pipes can withstand a maximum stress of 100 MPa. Both pipes have an outer radius of 1.8 cm. Pipe A has an inner radius of 1.5 cm, and Pipe B has an inner radius of 1 cm. The ratio of the maximum bending moment applied to two metallic pipes, each with a different inner and outer radius, is determined by considering their dimensions. The inner radius of the first pipe is 1.5 cm, and for...
191