Limitations of constant optimization with disclosed intraocular lens power formulae
Achim Langenbucher1, Peter Hoffmann, Alan Cayless
1From the Department of Experimental Ophthalmology, Saarland University, Homburg/Saar, Germany (Langenbucher, Wendelstein); Augen- und Laserklinik Castrop-Rauxel, Castrop-Rauxel, Germany (Hoffmann); School of Physical Sciences, The Open University, Milton Keynes, United Kingdom (Cayless); Department of Ophthalmology, Johannes Kepler University Linz, Linz, Austria (Wendelstein); Dr. Rolf M. Schwiete Center for Limbal Stem Cell and Aniridia Research, Saarland University, Homburg/Saar, Germany (Szentmáry); Department of Ophthalmology, Semmelweis-University, Budapest, Hungary (Szentmáry).
Purpose:
To investigate the effect of formula constants on predicted refraction and limitations of constant optimization for classical and modern intraocular lens (IOL) power calculation formulae.
Setting:
Tertiary care center.
Design:
Retrospective single-center consecutive case series.
Methods:
This analysis is based on a dataset of 888 eyes before and after cataract surgery with IOL implantation (Hoya Vivinex). Spherical equivalent refraction predSEQ was predicted using IOLMaster 700 data, IOL power, and formula constants from IOLCon ( https://iolcon.org ). The formula prediction error (PE) was derived as predSEQ minus achieved spherical equivalent refraction for the SRKT, Hoffer Q, Holladay, Haigis, and Castrop formulae. The gradient of predSEQ (gradSEQ) as a measure for the effect of the constants on refraction was calculated and used for constant optimization.
Results:
Using initial formula constants, the mean PE was -0.1782 ± 0.4450, -0.1814 ± 0.4159, -0.1702 ± 0.4207, -0.1211 ± 0.3740, and -0.1912 ± 0.3449 diopters (D) for the SRKT, Hoffer Q, Holladay, Haigis, and Castrop formulas, respectively. gradSEQ for all formula constants (except gradSEQ for the Castrop R) decay with axial length because of interaction with the effective lens position (ELP). Constant optimization for a zero mean PE (SD: 0.4410, 0.4307, 0.4272, 0.3742, 0.3436 D) results in a change in the PE trend over axial length in all formulae where the constant acts directly on the ELP.
Conclusions:
With IOL power calculation formulae where the constant(s) act directly on the ELP, a change in constant(s) always changes the trend of the PE according to gradSEQ. Formulae where at least 1 constant does not act on the ELP have more flexibility to zero the mean or median PE without coupling with a PE trend error over axial length.
More Related Videos
Related Concept Videos
Focusing of Light in the Eye
Glaucoma: Overview
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...