Dose optimization of 2D X-ray image acquisition protocols in image-guided radiotherapy
Marios Velonis1, Emmanouil Papanastasiou2, Konstantinos Hatziioannou3
1Department of Medicine, Faculty of Health Sciences, Democritus University of Thrace, Greece; Department of Medical Physics, Papageorgiou General Hospital, Thessaloniki, Greece.
Purpose:
In contemporary radiotherapy, patient positioning accuracy relies on kV imaging. This study aims at optimizing planar kV image acquisition protocols regarding patient dose without degrading image quality.
Materials And Methods:
An image quality test-object was placed in-between PMMA plates, suitably arranged to model head or pelvis. Constructed phantoms were imaged using default protocols, the resultant image quality was assessed and the corresponding radiation dose was measured. The process was repeated using numerous kV/mAs combinations to identify those acquisition settings providing images at lower dose than the default protocols but without deterioration in image quality. Default and dose-optimized protocols were then tested on an anthropomorphic phantom and on 51 patients during two successive treatment sessions. Image quality was independently assessed by two readers. Organ and effective doses were estimated using a Monte Carlo simulation software.
Results:
Low-contrast detectability exhibited a stronger dependence on kV/mAs settings, compared to high-contrast resolution. Dose-optimized protocols resulted in significant dose reductions (anteroposterior-head 48.0 %, lateral-head 30.0 %, anteroposterior-pelvis 28.4 %, lateral-pelvis 27.0 %) compared to the default ones, without compromising image quality. Optimized protocols decreased effective doses by 54 % and 29.6 % in head and pelvic acquisitions, respectively. Regarding image quality, anthropomorphic and patient images acquired using the dose-optimized protocols were subjectively evaluated equivalent to those obtained with the corresponding default settings, indicating that the proposed protocols may be routinely used.
Conclusions:
Given the potentially large number of radiotherapy fractions and the pertinent image acquisitions, dose-optimized protocols could significantly reduce patient dose associated with planar imaging without compromising positioning accuracy.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
X-ray Imaging