Radio Frequency Identification Gate System to Identify Misused Personal Dosimeters
Megumi Tamura1, Takeshi Kawamoto2, Minoru Ishifuro3,4
1Department of Clinical Support, Hiroshima University Hospital, Hiroshima, Japan. sawatamu@hiroshima-u.ac.jp.
A new radio frequency identification (RFID) gate system accurately identifies incorrectly positioned personal dosimeters, improving radiation dose estimation. This system alerts workers to misuse, enhancing safety and administrative efficiency in radiation monitoring.
Area of Science:
- Medical Physics
- Radiation Detection and Measurement
- Health and Safety Engineering
Background:
- Accurate personal dosimeter placement is crucial for effective dose estimation in radiation environments.
- Incorrect dosimeter positioning is common, leading to unreliable exposure records and administrative burdens.
- Existing methods for identifying mispositioned dosimeters are time-consuming and labor-intensive.
Purpose of the Study:
- To develop and evaluate a radio frequency identification (RFID) gate system for automatically identifying incorrectly positioned personal dosimeters.
- To differentiate between dosimeters worn over and under protective aprons.
- To provide real-time alerts for dosimeter misuse.
Main Methods:
- Development of an RFID gate system to read RFID-tagged dosimeters.
- Simulation of four dosimeter-wearing classes: proper use, reversed, both under, and both over.
- Evaluation of system performance using a confusion matrix.
- Assessment of system safety concerning life support devices.
Main Results:
- The RFID gate system achieved an overall accuracy of 97.75% in classifying dosimeter positions.
- The system successfully differentiated between dosimeters placed over and under a metal apron.
- No adverse effects on life support devices were observed at distances of 0.3 m or greater from the RFID antenna.
Conclusions:
- The developed RFID gate system is highly effective in identifying incorrectly positioned dosimeters.
- Real-time monitoring of dosimeter positioning can be achieved, improving radiation dose management.
- The system offers a practical solution to reduce administrative workload and enhance worker safety.