Characteristics of Option and Denali Inferior Vena Cava Filters
Lihao Qin1, Xiaocheng Gu1, Kai Wang1
1Department of Interventional and Vascular Surgery, The Affiliated Changzhou Second People's Hospital, Nanjing Medical University, Changzhou, China.
Option filters tilt more and change angle post-placement compared to Denali filters. Denali filters cause more IVC constriction, while Option filters require longer retrieval fluoroscopy times due to tilt.
Area of Science:
- Vascular surgery
- Medical device engineering
- Radiology
Background:
- Inferior Vena Cava (IVC) filters are crucial for preventing pulmonary embolism.
- Option and Denali filters are commonly used IVC filters with distinct designs.
- Understanding their performance characteristics is vital for clinical decision-making.
Purpose of the Study:
- To compare the placement and retrieval characteristics of Option and Denali IVC filters.
- To identify associations between filter characteristics and patient anatomy.
- To evaluate the incidence of filter tilt, tilt angle changes, and IVC constriction.
Main Methods:
- Retrospective study of 119 patients undergoing filter placement between August 2019 and August 2022.
- Inclusion of patients who received either Option or Denali filters.
- Analysis of filter tilt, tilt angle changes, retrieval success, IVC contraction, and fluoroscopy time.
Main Results:
- Option filters showed significantly higher tilt incidence (76.7%) and tilt angle change post-placement (74.1%) compared to Denali filters (11.9% and 8.3%, respectively).
- Option filter tilt at placement correlated with the common iliac vein-inferior vena cava (CIV-IVC) angle.
- Denali filters demonstrated a higher IVC contraction rate (87.5%) post-retrieval than Option filters (22.2%).
- Option filter retrieval required longer median fluoroscopy times, associated with tilt angle during retrieval.
Conclusions:
- Option filters exhibit greater propensity for tilt at placement and post-placement angle changes.
- The tilt angle of Option filters is influenced by the CIV-IVC angle.
- Option filter retrieval is associated with increased fluoroscopy time due to tilt.
- Denali filters are more likely to cause IVC constriction after placement.