Ultra-low-dose CT for attenuation correction: dose savings and effect on PET quantification for protocols with and
Natalie Anne Bebbington1, Kenneth Boye Christensen2, Lone Lange Østergård2
1Siemens Healthcare A/S, Borupvang 9, 2750, Ballerup, Denmark. natalie.bebbington@siemens-healthineers.com.
Background:
Ultra-low-dose (ULD) computed tomography (CT) scans should be used when CT is performed only for attenuation correction (AC) of positron emission tomography (PET) data. A tin filter can be used in addition to the standard aluminium bowtie filter to reduce CT radiation dose to patients. The aim was to determine how low CT doses can be, when utilised for PET AC, with and without the tin filter, whilst providing adequate PET quantification.
Methods:
A water-filled NEMA image quality phantom was imaged in three configurations with 18F-FDG: (1) water only (0HU); (2) with cylindrical insert containing homogenous mix of sand, flour and water (SFW, approximately 475HU); (3) with cylindrical insert containing sand (approximately 1100HU). Each underwent one-bed-position (26.3 cm) PET-CT comprising 1 PET and 13 CT acquisitions. CT acquisitions with tube current modulation were performed at 120 kV/50 mAs-ref (reference standard), 100 kV/7 mAs-ref (standard ULDCT for PET AC protocol), Sn140kV (mAs range 7-50-ref) and Sn100kV (mAs range 12-400-ref). PET data were reconstructed with μ-maps provided by each CT dataset, and PET activity concentration measured in each reconstruction. Differences in CT dose length product (DLP) and PET quantification were determined relative to the reference standard.
Results:
At each tube voltage, changes in PET quantification were greater with increasing density and reducing mAs. Compared with the reference standard, differences in PET quantification for the standard ULDCT protocol for the three phantoms were ≤ 1.7%, with the water phantom providing a DLP of 7mGy.cm. With tin filter at Sn100kV, differences in PET quantification were negligible (≤ 1.2%) for all phantoms down to 50mAs-ref, proving a DLP of 2.8mGy.cm, at 60% dose reduction compared with standard ULDCT protocol. Below 50mAs-ref, differences in PET quantification were > 2% for at least one phantom (2.3% at 25mAs-ref in SFW; 6.4% at 12mAs-ref in sand). At Sn140kV/7mAs-ref, quantification differences were ≤ 0.6% in water, giving 3.8mGy.cm DLP, but increased to > 2% at bone-equivalent densities.
Conclusions:
CT protocols for PET AC can provide ultra-low doses with adequate PET quantification. The tin filter can allow 60-87% lower dose than the standard ULDCT protocol for PET AC, depending on tissue density and accepted change in PET quantification.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography