Related Experiment Video
Updated: Jul 12, 2025

Robotic Ablation of Atrial Fibrillation
Published on: May 29, 2015
Comparative Effectiveness of Various Radiofrequency Ablation Catheters in the Ablation of Typical Atrial Flutter
Asher Gorantla1, Mahmoud Alsaiqali1, Jonathan Francois2
1Department of Medicine, SUNY Health Sciences University, Brooklyn, NY, USA.
Introduction:
Although ablation of typical atrial flutter (AFL) can be easily achieved with radiofrequency energy (RF), no studies compare the effectiveness of different ablation catheters. Our study aimed to compare the efficacy of various types of ablation catheters in treating typical AFL.
Methods:
We analyzed patients with AFL who underwent RF ablation by a single operator at our institution. Successful ablation was evidenced by a bidirectional conduction block (trans-isthmus conduction time ≥ 130 ms or double potentials ≥ 90 ms). Logistic regression was used to compare success rate and linear regression to compare lesion time.
Results:
Out of 222 patients, only six did not meet the success criteria (2.7%). The catheters used were non-irrigated, large-tip, internally irrigated (Chili II Boston Scientific), and externally irrigated (non-force-sensing) catheters (Cool Path, Abbott). An externally irrigated force-sensing catheter (TactiCath, Abbott) was used with > 10 gm of force and (LPLD) setting (30 W- 45 °C- 60 s), and high-power short-duration (HPSD) setting (50 W- 43 °C - 12 s). No complications were encountered. The catheter type had no statistically significant association with ablation success. With the use of externally irrigated catheter with contract force-sensing and HPSD settings, statistically significantly shortening of lesion time was achieved 758.3 s, [CI - 1128.29, - 388.35 s] followed by LPLD by 419.0 s [CI - 808.49, - 29.47 s].
Conclusions:
The typical atrial flutter radiofrequency ablation procedure had a high success rate, which was not influenced by the type of ablation catheter. Contact force ablation catheter and HPSD are associated with shorter total lesion time.
Related Concept Videos
Varicose Veins II: Diagnostic Studies and Interprofessional Care
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow...
Antiarrhythmic Drugs: Class IV Agents as Calcium Channel Blockers
Verapamil, a calcium channel blocker, inhibits calcium movement across myocardial cell membranes and vascular smooth muscle. This results in the dilation of coronary and...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias VI: Management of Dysrhythmias
Antiarrhythmic Drugs: Class II Agents as β-Adrenergic Blockers