Related Experiment Video
Updated: Jul 12, 2025

A Spine Robotic-Assisted Navigation System for Pedicle Screw Placement
Published on: May 11, 2020
Development and evaluation of robot-assisted ultrasound navigation system for pedicle screw placement: An ex-vivo
Ruixuan Li1, Ayoob Davoodi1, Yuyu Cai1
1Department of Mechanical Engineering, Robot-Assisted Surgery Group, KU Leuven, Leuven, Belgium.
Purpose:
Spinal instrumentation with pedicle screw placement (PSP) is an important surgical technique for spinal diseases. Accurate screw trajectory is a prerequisite for PSP. Ultrasound (US) imaging with robot-assisted system forms a non-radiative alternative to provide precise screw trajectory. This study reports on the development and assessment of US navigation for this application.
Methods:
A robot-assisted US reconstruction was proposed and an automatic CT-to-US registration algorithm was investigated, allowing the registration of screw trajectories. Experiments were conducted on ex-vivo lamb spines to evaluate system performance.
Results:
In total, 72 screw trajectories are measured, displaying an average position accuracy of 2.80 ± 1.14 mm and orientation accuracy of 1.38 ± 0.61°.
Conclusion:
The experimental results demonstrate the feasibility of proposed US system. This work, although restricted to laboratory settings, encourages further exploration of the potential of this technology in clinical practice.