Related Experiment Video
Updated: Jul 12, 2025

Radiotracer Administration for High Temporal Resolution Positron Emission Tomography of the Human Brain: Application to FDG-fPET
Published on: October 22, 2019
Miniature and flexible Bazooka balun for high-field MRI
1Department of Electrical and Computer Engineering, Vanderbilt University, Nashville, TN 37232, USA; Vanderbilt University Institute of Imaging Science, Vanderbilt University Medical Center, Nashville, TN 37232, USA.
Abstract:
Flexible coils offer improved patient comfort and better imaging quality. However, rigid and bulky baluns in RF coils limit flexibility and manufacturing. A miniaturized and flexible balun design was proposed to address this issue. It replaced rigid components with an ultra-flexible rubber tube and a flexible coaxial capacitor. Simulations validated the concept, and bench tests confirmed its performance, including a measured common-mode rejection ratio of -15.8 dB. The flexible balun was integrated into a 4-channel coil array, evaluating impedance changes caused by the "hand effect." Compared to coils without the balun, the flexible coil with the proposed balun showed improved robustness in impedance matching and inter-element couplings. Transmit efficiency of the flexible coil with the balun was compared to coils without a balun and with a rigid, shielded cable trap. Results demonstrated that the proposed balun circuit maintained high transmit efficiency. Overall, the flexible balun design offers a promising solution for improving the flexibility and performance of RF coil arrays in MRI applications.